What are the Krypton® DIS Flow Rate Requirements?
To produce a precise and consistent disinfectant measurement, a constant flow of water through the assembly is required. Learn about these requirements below.
The Kuntze Argon StabiFlow® regulates and controls the flow through system sample line at a constant and consistent 8 gallons per hour (GPH). To achieve this, the inlet flow rate needs to be 8.5 GPH or higher.
The Argon StabiFlow® also requires at least 5 psi of pressure, up to a maximum of 87 psi on the inlet.
Note: The Kuntze disinfectant measurement is pressure proof up to 87 psi, but the more controlled and lower the pressure, the less opportunity for operational hiccups.