Skip to Main Content

Cleaning in process

This article explains the Cleaning in Progress message shown on the analyzer during an active ASR® cleaning cycle.


Message Meaning

Cleaning in Progress is an event notification, not an alarm. It indicates that ASR® cleaning is currently active.

The message is typically displayed for 2 to 5 minutes.


System Behavior During Cleaning

  • Measurement is temporarily inactive during the cleaning cycle.
  • The last valid measured value is held (frozen) on the instrument display and mA output.
  • Control functions are temporarily disabled during cleaning.
  • Calibration is not possible while cleaning is active.

Calibration Recommendation

Do not calibrate immediately after ASR® cleaning.

Wait 15 to 30 minutes after cleaning has finished to allow the sensor signal to stabilize before starting calibration.


Recommended Cleaning Schedule

If possible, schedule ASR® cleaning for off-hours (for example around midnight) or for a time window when no operator is on site.

  • This reduces confusion caused by temporary “Cleaning in Progress” messages during active shifts.
  • It helps avoid calibration attempts during or directly after cleaning.
  • Keep the 15 to 30 minute stabilization period before calibration.

Summary

The Cleaning in Progress message confirms that ASR® cleaning is running normally. During this period, measurement and control are temporarily paused. After cleaning, allow sufficient stabilization time before calibration for reliable results.