Skip to Main Content

mA Output Add-On – Krypton® Multi

Learn about the mA output and how to set it for the Krypton® Multi

Helpful articles: How to get Add-On Codes – Krypton® Multi & Activate Add-Ons – Krypton® Multi

As soon as the mA outputs have been activated, the symbol mA OUT appears in the submenu IN/OUT.

Here you can set the following parameters: Main menu IN/OUT > mA OUT > choose an output > set parameters

Parameters:

  1. Assignment – Select which parameter you want to read out via the mA output: pH, DIS 1, optionally ORP, conductivity, DIS 2, or control variables
  2. Range – Select either 0-20 mA or 4-20 mA from the drop-down list. While 4-20 mA gives slightly less resolution, it helps identifying cable breaks

Note: Upon startup, the mA outputs are set to “not used” since they are equipped with a load monitor. If you assign a parameter to a mA output while the output is disconnected or the connection is interrupted somewhere, the event message “load impedance mA” appears in the display

  1. 22 mA alarm – If you activate this function, the mA output rises to 22 mA during an active alarm.
  2. Range min and max – You can zoom by defining which measured values correspond to 0/4 mA and which to 20 mA. With these settings you can even invert the output, by assigning 20 mA to a smaller value than 0/4 mA. In case of controller output, the output is in %, so you cannot set range min and max

Note: If the output is assigned to control, the 22 mA alarm is deactivated. It needs to be reactivated if the output is reassigned to measurement.