Lesson 1 – Calibrating Your Zirkon® pH Sensor (4:05)

Watch the video to learn how to calibrate your Zirkon® pH sensor.

Best Practices pH Calibration

  1. Input calibration settings.

Menu > Cal > Cal pH > Settings

  1. Remove temperature and pH sensors.
  2. Navigate to pH calibration menu.

Menu > Cal > Cal pH > Run

  1. Place pH sensor and temperature sensor in buffer solution and press “start.”
  2. Rinse both sensors with DI water.
  3. Place pH sensor and temperature sensor in buffer solution and press “start.”
  4. Read slope and either repeat process or press “Save.”
  5. Reinstall sensors and restore flow.