Lesson 2 – Calibrating Your Zirkon® DIS Sensor (2:44)
Watch the video to learn how to calibrate your Zirkon® DIS sensor.
Best Practices DIS Calibration
- Manually clean and install sensor.
- Restore flow through the flow cell.
- Navigate to DIS calibration menu.
Menu > Cal > Cal DIS 1 > Reference
- Take sample(s) from sample valve and take reference measurement(s).
- Enter reference measurement result and press “OK.”
- Read slope and either repeat process or press “Save.”